[U(mV)-T(°C) calibration]
FitMethod = Polynom
TransformX = Line
TransformY = CALC_TC_B
Power = 1
Center = 0
Scale = 1
Fixed = 1
Coeff[0] = 0
Coeff[1] = 1
Bounds = 0.000 13.820
Data U(mV)  T(°C)  Weight Tcj(°C)
     0.000  0      1      0
     0.033  100    1      0
     0.178  200    1      0
     0.431  300    1      0
     0.787  400    1      0
     1.242  500    1      0
     1.792  600    1      0
     2.431  700    1      0
     3.154  800    1      0
     3.957  900    1      0
     4.834  1000   1      0
     5.780  1100   1      0
     6.786  1200   1      0
     7.848  1300   1      0
     8.956  1400   1      0
     10.099 1500   1      0
     11.263 1600   1      0
     12.433 1700   1      0
     13.591 1800   1      0
     13.820 1820   1      0
End Data
Notice Text
Standard calibration of thermocouple type B (Pt-30%Rh/Pt-6%Rh) ТПР платина-родий platinum-rhodium
Reference source: GOST R 8.585—2001, Table 3; IEC 60584-1(1995)
Configuration example:
Link AnalogInput  0 with curve ThermoCoupleVoltage(mV)
Link AnalogInput  1 with curve TemperatureColdJunction(°C)
Link AnalogOutput 0 with curve TemperatureThermoCouple(°C) history 100
Calibration#0 = ~~\Resource\DaqSite\StdLib\Calibr\_CALC_TC_B.CAL U(mV) T(°C) Tcj(°C) Line CALC_TC_B 0.000 13.820