[R(OHM)-T(°C) calibration]
FitMethod = Polynom
TransformX = Line
TransformY = CALC_RTD_CU100_426
Power = 1
Center = 0
Scale = 1
Fixed = 1
Coeff[0] = 0
Coeff[1] = 1
Bounds = 78.7 185.2
Data R(OHM) T(°C)  Weight
     78.7   -50    1     
     82.96  -40    1     
     87.22  -30    1     
     91.48  -20    1     
     95.74  -10    1     
     100    0      1     
     108.52 20     1     
     117.04 40     1     
     125.56 60     1     
     134.08 80     1     
     142.6  100    1     
     151.12 120    1     
     159.64 140    1     
     168.16 160    1     
     176.68 180    1     
     185.2  200    1     
End Data
Notice Text
Standard calibration of RTD Cu100 copper медь R0=100 alfa=0.00426
Reference source: GOST 6651—2009, Table A4
Configuration example:
Link AnalogInput  0 with curve RTDResistance(OHM)
Link AnalogOutput 0 with curve RTDTemperature(°C) history 100
Calibration#0 = ~~\Resource\DaqSite\StdLib\Calibr\_CALC_RTD_CU100_426.CAL R(OHM) T(°C) * Line CALC_RTD_CU100_426 78.7 185.2
End Notice Text