[U(mV)-T(°C) calibration]
FitMethod = Polynom
TransformX = Line
TransformY = GOST_TC_S
Power = 1
Center = 0
Scale = 1
Fixed = 1
Coeff[0] = 0
Coeff[1] = 1
Bounds = -0.236 18.693
Data U(mV)  T(°C)  Weight Tcj(°C)
     -0.236 -50    1      0
     -0.194 -40    1      0
     -0.150 -30    1      0
     -0.103 -20    1      0
     -0.053 -10    1      0
     0.000  0      1      0
     0.646  100    1      0
     1.441  200    1      0
     2.323  300    1      0
     3.259  400    1      0
     4.233  500    1      0
     5.239  600    1      0
     6.275  700    1      0
     7.345  800    1      0
     8.449  900    1      0
     9.587  1000   1      0
     10.757 1100   1      0
     11.951 1200   1      0
     13.159 1300   1      0
     14.373 1400   1      0
     15.582 1500   1      0
     16.777 1600   1      0
     17.947 1700   1      0
     18.693 1768   1      0
End Data
Notice Text
Standard calibration of thermocouple type S (Pt-10%Rh/Pt) ТПП10 платина-родий/платина platinum-rhodium/platinum
Reference source: GOST R 8.585—2001, Table 2; IEC 60584-1(1995)
Configuration example:
Link AnalogInput  0 with curve ThermoCoupleVoltage(mV)
Link AnalogInput  1 with curve TemperatureColdJunction(°C)
Link AnalogOutput 0 with curve TemperatureThermoCouple(°C) history 100
Calibration#0 = ~~\Resource\DaqSite\StdLib\Calibr\_GOST_TC_S.CAL U(mV) T(°C) Tcj(°C) Line GOST_TC_S -0.236 18.693
End Notice Text