[U(mV)-T(°C) calibration]
FitMethod = Polynom
TransformX = Line
TransformY = CALC_TC_N
Power = 1
Center = 0
Scale = 1
Fixed = 1
Coeff[0] = 0
Coeff[1] = 1
Bounds = -4.345 47.513
Data U(mV)  T(°C)  Weight Tcj(°C)
     -4.345 -270   1      0
     -4.313 -250   1      0
     -4.196 -225   1      0
     -3.990 -200   1      0
     -3.702 -175   1      0
     -3.336 -150   1      0
     -2.902 -125   1      0
     -2.407 -100   1      0
     -1.859 -75    1      0
     -1.269 -50    1      0
     -0.646 -25    1      0
     0.000  0      1      0
     2.774  100    1      0
     5.913  200    1      0
     9.341  300    1      0
     12.974 400    1      0
     16.748 500    1      0
     20.613 600    1      0
     24.527 700    1      0
     28.455 800    1      0
     32.371 900    1      0
     36.256 1000   1      0
     40.087 1100   1      0
     43.846 1200   1      0
     47.513 1300   1      0
End Data
Notice Text
Standard calibration of thermocouple type N (Ni-Cr-Si/Ni-Si) ТНН нихросил/нисил nicrosil/nisil
Reference source: GOST R 8.585—2001, Table 8; IEC 60584-1(1995)
Configuration example:
Link AnalogInput  0 with curve ThermoCoupleVoltage(mV)
Link AnalogInput  1 with curve TemperatureColdJunction(°C)
Link AnalogOutput 0 with curve TemperatureThermoCouple(°C) history 100
Calibration#0 = ~~\Resource\DaqSite\StdLib\Calibr\_CALC_TC_N.CAL U(mV) T(°C) Tcj(°C) Line CALC_TC_N -4.345 47.513
End Notice Text