[U(mV)-T(°C) calibration]
FitMethod = Polynom
TransformX = Line
TransformY = GOST_TC_R
Power = 1
Center = 0
Scale = 1
Fixed = 1
Coeff[0] = 0
Coeff[1] = 1
Bounds = -0.226 21.101
Data U(mV)  T(°C)  Weight Tcj(°C)
     -0.226 -50    1      0
     -0.188 -40    1      0
     -0.145 -30    1      0
     -0.100 -20    1      0
     -0.051 -10    1      0
     0.000  0      1      0
     0.647  100    1      0
     1.469  200    1      0
     2.401  300    1      0
     3.408  400    1      0
     4.471  500    1      0
     5.583  600    1      0
     6.743  700    1      0
     7.950  800    1      0
     9.205  900    1      0
     10.506 1000   1      0
     11.850 1100   1      0
     13.228 1200   1      0
     14.629 1300   1      0
     16.040 1400   1      0
     17.451 1500   1      0
     18.849 1600   1      0
     20.222 1700   1      0
     21.101 1768   1      0
End Data
Notice Text
Standard calibration of thermocouple type R (Pt-13%Rh/Pt) ТПП13 платина-родий/платина platinum-rhodium/platinum
Reference source: GOST R 8.585—2001, Table 1; IEC 60584-1(1995)
Configuration example:
Link AnalogInput  0 with curve ThermoCoupleVoltage(mV)
Link AnalogInput  1 with curve TemperatureColdJunction(°C)
Link AnalogOutput 0 with curve TemperatureThermoCouple(°C) history 100
Calibration#0 = ~~\Resource\DaqSite\StdLib\Calibr\_GOST_TC_R.CAL U(mV) T(°C) Tcj(°C) Line GOST_TC_R -0.226 21.101
End Notice Text